In an effort to meet the needs of calibration facilities which offer NIST
traceable calibrations of air kerma measuring instruments used in the
mammography energy range, NIST will offer proficiency tests. The FDA-MQSA
"
Final
Rule for Quality Mammography Standards"
requires that all
clinical
mammography units be calibrated with an instrument
that is traceable
to a
national standard. The facility that provides the NIST
traceable
calibration
must participate in a biennial proficiency program
directly with
NIST. The
calibration facility must show a 3 % agreement
with the
national standard
in the mammography energy range.
Proficiency Tests for Mammography
- The calibration facility shall initiate contact with NIST and request a
proficiency test for mammography calibrations.
- NIST will conduct a proficiency test on an individual basis, as well as
part of an accreditation program.
- The proficiency test will consist of three stages. First NIST will
calibrate a NIST transfer standard using two appropriate molybdenum and/or
rhodium generated x-ray beam qualities. NIST will then send the transfer
standard to the participating calibration facility, along with details of the
NIST calibration conditions, excluding the calibration results. Calibrations of
the NIST transfer chamber will be made at the facility using the appropriate
beam qualities. The NIST transfer chamber shall be returned to NIST
accompanied with the calibration results in the form of Gy/C, for each beam
quality. Upon receipt of the chamber at NIST, it will be re-calibrated to
assure stability and proper working conditions.
- The results, provided by the participating facility, will be compared to
the NIST calibrations. A NIST proficiency report will be issued to each
participating calibration facility. The report will state the level of
agreement with NIST under the conditions of the proficiency test. The
proficiency test can be used to demonstrate that the participating facility
can calibrate a NIST transfer standard, using techniques, characterized x-rays
and NIST traceable instruments of sufficient quality permitting acceptable
agreement with NIST. It will be the responsibility of the calibration facility
to inform the FDA-MQSA of the proficiency test results.
- The NIST mammography calibration facility will be used for all mammography
calibrations and proficiency tests. It is the opinion of NIST that the
molybdenum and rhodium generated x-ray beam qualities best represent the end
users’ calibration conditions. Since medical physicists and inspectors are
required to use NIST traceable instruments to conduct inspections of clinical
molybdenum and/or rhodium anode x-ray units, the calibrations of the
mammography calibration instruments will be conducted using two of the
seventeen NIST Mo and/or Rh beam qualities. The selection of the two beam
qualities to be used in the proficiency test will be made jointly by NIST and
the participating calibration facility. The beam qualities chosen will be
within the HVL and energy range of the beams available for calibrations at the
participating facility. The facility may choose to compare with NIST for more
than the two required beam qualities. Information concerning the air kerma
rate, calibration distance and beam size used for the NIST calibrations will be
issued to the participating calibration facility at the time of receipt of the
NIST transfer standard. However, the calibration facility should use their
routine calibration conditions for proficiency tests. The goal of the
proficiency test is to test that the calibration facility is transferring
quality calibrations under routine conditions.
- NIST will conduct the proficiency test using a NIST-owned transfer-quality
ionization chamber. The participating facility is responsible for obtaining the
necessary electrometer. NIST will not provide a chamber/electrometer unit.
- The proficiency test, which will consist of a test using a minimum of two
beam qualities, will cost the same as two NIST calibrations plus
15 percent. If the calibration facility elects to be tested at more than
two beam qualities, the cost will increase.
Gallery
| Mammography
| Radiation Interactions and Dosimetry Group
| Ionizing Radiation Division
Inquiries or comments:
michelle.obrien@nist.gov
Online: November 1998 - Last update: March 2005
|